Apollo
Safety Features
10.2 Mains Power Failure
If the battery is almost depleted, the message “BATTERY LOW
will be displayed. Apollo permits manual ventilation with 100% O2
in the event of a power failure and depleted battery. The fresh gas
measurements, ventilator, and monitoring are inactive.
In case of complete power failure:
Close N2O and AIR flow valves
Check vaporizer settings
Open O2 flow to the desired level (using total flow meter)
Ventilate patient manually