Apollo
Safety Features
10.2 Mains Power Failure
In the event of power failure the Apollo automatically switches to
the built-in uninterruptible power supply (battery backup). Provided
that the battery is fully charged, operation will be continued for at
least 30 minutes (up to 90 minutes, depending on the ventilation
parameters).
The message “POWER FAIL” is displayed in the status field on the screen,
together with the remaining battery capacity in percent.