Supplement | Atlan A100 (XL), A300 (XL), A350 (XL) 5
Supplement to the instructions for use | English
Atlan
A100 (XL), A300 (XL), A350 (XL)
Supplement to the instructions for use
Keep this supplement with the instructions for use.
This supplement updates the information of the instructions for use in the following
chapters.
Assembly and preparation
Safety instructions
Electrical safety
The following section has been changed:
Battery supply
If the battery is not sufficiently charged, it may not be possible to maintain operation
for the required period if the mains power supply fails. As a result, the patient could
be put at risk.
Before first operation or after storage, charge the battery for at least 8 hours.
Avoid short charge and discharge cycles.
When the device is switched on, do not disconnect it from the mains power
supply.
When the device is powered by the internal battery while switched on, ensure
the functional integrity of the battery by performing a battery test at regular
intervals. For further information see: "Inspection", page 6.
Instructions for use Part number Edition
Atlan A300, A300 XL, A350,
A350 XL SW 1.0n
9056001 1 2019-02 and higher
Atlan A300, A300 XL, A350,
A350 XL SW 2.0n
9510601 1 2021-06 and higher
Atlan A100 (XL), A300 (XL),
A350 (XL) SW 2.1n
9511481 1 2022-08 and higher

Other manuals for Dräger Atlan A300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A300 and is the answer not in the manual?

Dräger Atlan A300 Specifications

General IconGeneral
BrandDräger
ModelAtlan A300
CategoryMedical Equipment
LanguageEnglish

Related product manuals