CAUTION !
Restriction of Distribution
Federal Law and Regulations in the United
States and Canada restrict this device to sale by
or on the order of a physician.
WARNING !
If a fault is detected in the ventilator and its
life-support functions are in doubt, ventilation
must be started without delay with an inde-
pendent ventilation device (resuscitation bag)
- using PEEP and/or increased inspiratory O2
concentration where necessary and appro-
priate. The unit should then be removed from
use and serviced by an authorized service
technician.
WARNING !
In case of malfunction of any of the built-in
monitoring a substitute must be provided in
order to maintain an adequate level of moni-
toring. The operator of the ventilator must still
assume full responsibility for proper venti-
lation and patient safety in all situations.
WARNING !
Always use independent apnea monitor when
the built-in apnea monitoring has been
switched off. The operator of the ventilator
must still assume full responsibility for proper
ventilation and patient safety in all situations.
WARNING !
Whenever a patient is connected to the venti-
lator, constant attention by qualified medical
staff is required in order to provide immediate
corrective action in case of a malfunction.
WARNING !
Sidestream monitors may cause negative
pressures in the event of a blocked inspiratory
circuit. Always connect monitoring lines of a
suctioning sidestream monitor via adapter
84 12 448 with safety valve.
WARNING !
Do not use ventilator in conjunction with
nuclear spin tomography (MRT, NMR, or NMI)!
Equipment malfunction may result.
WARNING !
Mobile telephones must not be used within 10
meters (33 feet) of the equipment. Mobile
telephones can interfere with the function of
electromedical equipment and therefore
endanger the patient!
Important Safety Information
Summary of WARNINGS and CAUTIONS
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.