Instructions for use Babylog 8000 plus SW 5.n 107
Troubleshooting
Calibrate flow sensor! The flow sensor must be cali-
brated after the device is
switched on and after a mains
power supply failure.
Press the OK key and calibrate
the flow sensor, see page 95.
The flow is not measured until
the sensor has been calibrated.
To work without flow measure-
ment, press only the OK key.
Check PEEP setting! The PEEP/CPAP setting knob
determines the mean airway
pressure during high-frequency
ventilation.
After high-frequency ventilation is
terminated, set PEEP to the
desired value using the
PEEP/CPAP setting knob.
Fault in rotary knob Device faulty. Contact DrägerService.
FiO
2 high FiO2 measurement faulty. Calibrate O2 measurement man-
ually, see page 94
Mixer function impaired. Contact DrägerService.
Air supply pressure too low. Ensure pressure greater than
2.7 bar (39.2 psi).
FiO
2 low FiO2 measurement faulty. Calibrate O2 measurement man-
ually, see page 94.
Mixer function impaired. Contact DrägerService.
O
2 supply pressure too low. Ensure pressure greater than
2.7 bar (39.2 psi).
Flow measurement disturbed
Measurement switched off
Flow sensor disconnected. Connect flow sensor or flow sen-
sor cable. Calibrate the flow sen-
sor, see page 95.
Flow sensor faulty. Replace the flow sensor, see
page 42.
Flow sensor cable faulty. Replace the flow sensor cable.
Flow measurement disturbed
VG uses Pinsp,
Check set value!
Volume regulation VG inter-
rupted because the flow sensor
is faulty or disconnected. Flow
measurement fault.
Replace flow sensor.
Connect flow sensor cable.Set
Pinsp to an appropriate setting.
Flow sensor dirty?
Please clean sensor!
Water or secretion in the flow
sensor.
Replace the flow sensor insert.
Flow sensor cable faulty. Replace the flow sensor cable.
Frequency high! Hyperventilation Set respiratory rate.
Auto-triggering Increase trigger threshold.
Hose kinked? Breathing hose kinked, blocked,
or condensate in breathing hose.
Check breathing hoses, establish
clear passage.
Inner diameter of breathing
hoses too small.
Use a suitable breathing circuit.