For your safety and that of your patients
14 Instructions for use Babylog 8000 plus SW 5.n
Ensuring ventilation using an
independent manual ventilator
CAUTION
Risk of patient injury
Do not make therapeutic decisions based solely
on individual measured values and monitoring
parameters.
CAUTION
Risk of patient injury
Positive-pressure ventilation can lead to negative
effects, such as barotrauma or strain on the
circulatory system.
Monitor the patient's condition.
CAUTION
Reduced O2 concentration
When nitric oxide (NO) is added, the set
inspiratory O2 concentration may be reduced.
Monitor the O
2 concentration.
CAUTION
Risk of medical device overheating
The medical device may overheat due to sources
of heat such as direct sunlight, heat radiators, or
spotlights.
Keep the medical device away from sources
of heat.
Only operate the medical device in rooms
with sufficient ventilation.
Do not cover the medical device or slide the
rear of the device against a wall.
CAUTION
Risk of medical device overheating
If the ventilation slots on the medical device are
covered or sealed, the medical device may
overheat.
Air must be able to enter freely.
An alarm is triggered if the medical device
overheats during operation.
CAUTION
Risk of personal injury
If the glass on the LC screen is damaged, a
chemical liquid may escape.
Avoid contact with the body.
Immediately clean affected skin areas with
soap.
WARNING
Risk of patient injury
If a fault is detected in the medical device, its
life-support functions may no longer be
assured.
Ventilation of the patient using an
independent ventilation device must be
started without delay, if necessary with PEEP
and/or an increased inspiratory O
2
concentration (e.g., with a manual
resuscitator).