EasyManua.ls Logo

Dräger Babylog 8000 plus - Page 151

Dräger Babylog 8000 plus
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Babylog 8000 plus SW 5.n 151
Technical data
Connection to IT networks
Data can be exchanged in an IT network using
wired and wireless technologies. IT networks
include all data interfaces (e.g., RS232, LAN, USB,
printer interface) described in standards and
conventions.
Information for connection to an IT
network
Preconditions
The device must only be connected to the network
by service personnel. The head of IT at the
hospital must be consulted in advance.
The following documents must be observed:
Documentation accompanying this device
Description of the network interface
Description of network-based alarm systems
Dräger recommends observing IEC 80001-1 (risk
management for IT networks with medical
devices).
Serial interfaces
The following interfaces are possible:
RS232 interfaces according to EIA RS-232
(CCITT V.24/V.28) for the following
applications:
–MEDIBUS
Connections with third-party medical
devices
Electrical requirements for connected devices
and networks
The serial interface is only appropriate for
connecting devices or networks that have a
nominal voltage on the network side of max.
24 V DC and meet the requirements of one of the
following standards:
IEC 60950-1: Ungrounded SELV electrical
circuits
IEC 60601-1 (2nd edition or later): Exposed
secondary circuits

Other manuals for Dräger Babylog 8000 plus

Related product manuals