Operating concept
30 Instructions for use Babylog 8000 plus SW 5.n
Exceeding setting limits
If the setting limit of a ventilation parameter is
exceeded, the LED for that setting knob flashes
and a message is displayed on the screen.
Acknowledge the message by pressing the OK
key and set the ventilation parameter with the
setting knob.
Setting PEEP < 2.5 mbar (2.5 cmH
2O)
Pull on the PEEP/CPAP setting knob and
simultaneously turn it to the left to the range
marked in red. Set the value for PEEP.
Keys with fixed and variable functions
Setting the ventilation mode
1 Press the Vent. Mode key (C).
2 Press the key (E) for the desired ventilation
mode, e.g., SIMV.
3 Press the On key (E).
The ventilation mode is active.
20925
22211
A Suppresses the acoustic alarm
signal for 2 minutes
B OK Acknowledges messages and
alarm messages
C Vent.
Mode
Opens the ventilation mode dia-
log window
D Vent.
Option
Opens the additional settings dia-
log window
E 6 variable
keys
Selects functions or settings,
changes ventilation modes
F Cal. Con-
fig.
Opens the calibration and config-
uration dialog window
G man. Insp. Triggers manual inspiration, see
page 69