Getting started
60 Instructions for use Babylog 8000 plus SW 5.n
2 Kink the expiratory breathing hose.
The visual alarm signal flashes red, the acoustic
alarm signal sounds, and one of the following
alarm messages is displayed:
3 Release the expiratory breathing hose.
4 Disconnect the connector on the Y-piece.
The visual alarm signal flashes red, the acoustic
alarm signal sounds, and one of the following
alarm messages is displayed:
5 Set the PEEP/CPAP setting knob back to 0.
6 Reconnect the test lung.
Checking the breathing gas humidifier
Check the breathing gas humidifier in
accordance with the corresponding instructions
for use.
Device check immediately before using
the device on a patient
Carry out the device check immediately before
using the device on a patient in order to check the
operability of the device.
Conditions:
Gas supply must be ensured.
The expiration valve, breathing circuit, and flow
sensor must be correctly connected.
The test lung must be connected.
Checking the breathing circuit for leaks
1 Switch the device on.
2 Press the OK key. The alarm message
Calibrate flow sensor! is no longer displayed.
3 Set the following ventilation parameters:
4 Press the OK key.
5 Press and hold the man. Insp. key. The bar
display must read (80 ±2) mbar.
Calibrating the flow sensor
Calibrate the flow sensor, see page 95.
If calibration is successful, a confirmation is
displayed on the screen.
Checking the ventilation functions
1 Set the following alarm limits:
Hose kinked?
Airway pressure high
Inspiration cancelled
Airway pressure low
Leak in hose system?
Check setting!
Parameter Setting
Ventilation mode CPAP
2
Pinsp 80
MV 0L/min
MV 15 L/min