88 Instructions for use Babylog 8000 plus SW 5.n
Displaying lung-mechanics parameters
The patient's resistance and compliance are
calculated using the linear-regression method.
Information on calculations can be found on
page 167.
After a manual inspiration, the display of
waveforms is automatically stopped for one
minute. The waveforms and associated measured
values can be evaluated together.
1 On the main screen, press the Meas key.
2 Press the RC key.
Displaying measured volume values
1 On the main screen, press the Meas key.
2 Press the Vol key.
The following measured volume values are
displayed (example):
Displaying combinations of measured values
1 On the main screen, press the Meas key.
2 Press the MV O
2 P key.
The following measured values are displayed
(example):
21075
C Dynamic compliance of the respiratory
system
R Airway resistance including the tube
C
20/C Index indicating if the lungs are being
overinflated. If C
20/C <0.8 the lungs
may be overinflated. Can only be used
with ventilation modes without plateau.
Tc Time constant of the respiratory sys-
tem in milliseconds
r Correlation coefficient of linear regres-
sion
If the warning symbol (A) is displayed
next to the value for r, the measured
values may be distorted, e.g., due to a
leakage.
A
21076
MV Expiratory minute volume
spont Percentage of spontaneous breathing
to minute volume
Leak Tube leakage
VT Expiratory tidal volume of the previous
respiratory cycle
21081
MV Expiratory minute volume
FiO
2 Measured inspiratory O2 concentration
Mean Mean airway pressure of previous
respiratory cycle