Instructions for use Babylog 8000 plus SW 5.n 97
Monitoring
Deactivating flow monitoring
Flow monitoring is deactivated when the flow
sensor cable is removed from the flow sensor.
1 Remove the plug (A) of the flow sensor cable
from the flow sensor (B).
The device displays an alarm message (C). An
information (D) is displayed in the status field.
2 Acknowledge the alarm message by pressing
the OK key.
Activating flow monitoring
1 Connect the plug (A) of the flow sensor cable to
the flow sensor.
2 Check whether flow monitoring is active. The
information (D) is no longer displayed in the
status field.