Supplement Fabius MRI SW 3.n 5
English
LEDs for alarm status
Fabius MRI has alarm LEDs and alarm LED bars
which signal alarms and indicate the alarm priority.
Further information can be found in the "Alarms"
chapter.
Alarm LEDs
Warning: Red LED (A) flashes
Caution: Yellow LED (B) flashes
Note: Yellow LED (B) lit constantly
Alarm LED bars
The alarm LED bars are triggered by the same
causes as the alarm LEDs (A and B).
Signaling in the event of a device failure
Example: Left-hand side of the device
If the mains power supply and the battery fail,
the device will no longer be ready for operation.
Only the red LEDs (D) in the alarm LED bars
(C) will be flashing.
2347623477
A
B
A
B
23455
CAUTION
Risk of alarms not being noticed
If the device is not positioned correctly, there is a
risk that alarms may be noticed too late or not at
all.
Position the device so that the alarm LED
bars are visible from various viewing angles
and from every location.
Do not cover the alarm LED bars.