Overview
10 Supplement Fabius plus SW 3.n
Ceiling-mounted version (optional)
The "Caution - Risk of tipping over" safety information has been
omitted and the following warning has been added in its place:
WARNING
Risk of tipping over
The ceiling-mounted version of the device
does not meet the requirements of ISO 80601-
2-13 and IEC 60601-1 with regard to tipping
stability and transport stability. Precautionary
measures must be taken. The medical device
may tip over if handled incorrectly.
The medical device may only be moved by
people who have the physical ability to do
so.
Use 2 persons to transport the device.
Make sure that there is sufficient clear
space.
Remove any devices mounted on attached
support arms or on the top of the device.
Fold in the support arms.
Clear the writing tray and push it in
completely.