EasyManua.ls Logo

Dräger Fabius plus - Breathing System

Dräger Fabius plus
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Supplement Fabius plus SW 3.n 45
Technical data
Breathing system
The following data have been added:
Alarm for low oxygen supply pressure
The alarm limit has changed:
Compliance
including ventilator hose (without breathing
hoses)
0.8 mL/cmH
2O (0.8 mL/hPa) at 30 hPa
Flexible arm for breathing bag (optional)
Volume 0.13 L
Compliance 0.13 mL/cmH
2O (0.13 mL/hPa) at 30 hPa
Rigid arm for breathing bag (optional)
Volume 0.11 L
Compliance 0.11 mL/cmH
2O (0.11 mL/hPa) at 30 hPa
Resistance
Peak flow
[L/min]
Resistance
[cmH
2O (hPa)]
Inspiratory Expiratory
In accordance with ISO 80601-2-13, dry, with
adult breathing hose set M30146
1)
1) Depending on the ventilation settings, the indicated values may deviate by ±0.3 cmH2O (0.3 hPa).
60 -4.7 4.4
30 -0.6 4.2
15 0.4 3.4
In accordance with ISO 80601-2-13, dry, with neo-
natal breathing hose set M33681
1)
2.5 1.0 2.9
In accordance with ISO 80601-2-13, dry, without
hoses
1)
60 -3.7 3.7
Alarm limit Warning signal (continuous tone for 10 seconds,
adjustable from approx. 61 dB(A) to 74 dB(A)), as
soon as the pressure drops below 20 ±6 psi (1.35
±0.4 kPa x 100).