23-16 DELTA/DELTA XL/KAPPA VF8
23 S
CIO® FOUR MODULES
Quick Reference Table -- O2/N2O Setup
Click on the following items to execute O
2
/N
2
O setup functions
The O
2
(/N
2
O) Setup Menu
Menu Item Description Settings
MultiGas
Parameter
Enables the combined MultiGas
parameter box display (see page
23-22).
• ON, OFF
O2 Scale Sets O
2
waveform scale • 50 %, 100 %
N2O Display Displays N
2
O values
NOTES: Ghosted if the monitor is
displaying a combined MultiGas
parameter box (see page 23-22).
N
2
O alarms are not supported, so N
2
O
alarm limits do not appear in the
parameter box.
•ON
•OFF
MultiGas Zero Manually zeroes the Scio module.
NOTE: During zeroing, the monitor
temporarily blanks Scio parameter
values.
Not applicable
Auto Zero Delay Delays automatic zeroing for 5 minutes
for uninterrupted monitoring (see page
23-13 for information)
Not applicable
O2 Calibration See page 23-12 for further information
on calibration functions.
• 1 Point Cal.
• 2 Point Cal.
• Last O2 Cal.
O2 Alarms Accesses the Limits table
(see chapter 5)
Not applicable