BIS SETUP
VF8 DELTA/DELTA XL/KAPPA 25-7
Show all
parameters
(continued)
PWR:Total Power
(Range = 40 - 100 dB in the
frequency range from 0.5 -
30 Hz)
A measure of absolute total
power (in decibels) in the
frequency range from 0.5 -
30 Hz.
BCT: Burst Count
(Range = 0 - 30)
An alternative to BSR for
quantifying the amount of
suppression in the EEG. It
gives the number of EEG
bursts in the last minute.
Burst Count is activated only
when an Extend sensor is
connected. BCT value is
blank for any other type of
sensor.
Check
impedance
Performs an impedance check (tests the
quality of BIS electrodes’ connection) of
the sensor (for more information, see
page 25-8)
Not applicable
BIS smoothing
rate
Selects the smoothing rate at which BIS
value is averaged
• 15 sec. (default): More responsive
to state changes (for example,
induction or awakening).
• 30 sec: A smoother trend (lower
variability and less sensitive to
artifact).
Filter Allows you to set filters to ON or OFF
NOTE:
The BISx pod uses filters to screen out
undesirable interference from the raw
EEG signal. If you prefer to observe the
raw EEG without the filters, they may be
turned OFF.
• ON (default)
•OFF
Parameter 2 Allows you to set second BIS parameter •BCT
•PWR
• SEF
• BSR (default)
SRS # of
Use(s)
available
Allows you to display the number (0-99) of
uses that remain available.
NOTE: This display is read only.
• Not applicable
Menu Item Description Available Settings