PICCO PARAMETER SETUP
VF8 DELTA/DELTA XL/KAPPA 26-21
Catheter type
NOTE: Other is
displayed if the
catheter that is
connected requires
manual entry of an
ACC value.
Selecting Other
displays a key pad
for manual entry.
Read-only field. Displays the
catheter type in use.
PV2013L07, PV2014L08,
PV2014L16, PV2014L22,
PV2015L20, PV2014L50, Other
Catheter Position Selects the position of the catheter.
NOTE: The selection Radial artery
is only visible when the catheter
PV2014L50 is connected.
Femoral artery, Axillary artery,
Brachial artery, Radial artery
NOTE: When a patient is initially
admitted, this field displays ‘---’ until
the catheter position is selected.
CAUTION: Confirm that the
selected catheter position is
accurate before starting a
measurement. Changing the
catheter position after starting a
measurement, has no effect on the
current measurements. To correct
the catheter position, discharge and
readmit the patient.
Art. Catheter
Const. (ACC)
Field ghosted when catheters with
automatic ACC detection are
connected. This field will be
unghosted and available to catheters
(other than Catheter Type field)
requiring manual entry of catheter
constant values (available range is
0-999). Selecting this key displays a
keypad for manual entry.
Not applicable
Manual CVP Allows you to enable/disable manual
CVP Value menu.
NOTE: If Manual CVP is Enabled,
CVP Value selection is ghosted
(unavailable).
Disabled (default), Enabled
Thermodilution Setup Menu