MONITORING S PECIFICATIONS
VF8 DELTA/DELTA XL/KAPPA B-35
ST-Segment Analysis
EEG
Sensed leads: 3 lead cable:I, II, or III (user-selectable)
5 lead cable:(choice of 3 leads) I, II, III, aVR, aVL, aVF, or V
6 lead cable:(choice of 3 leads) I, II, III, aVR, aVL,
(w/ TruST Off )aVF, V, V+
6 lead cable:(choice of 3 leads) I, II, III, aVR, aVL,
(w/ TruST On) aVF, V1-V6, dV1, dV3, dV4, dV6
NOTE: TruST leads are indicated by a prefix “d” before the
V lead.
12 lead cable: (choice of 3 leads) I, II, III, aVR, aVL,
(Optional) aVF, V1-V6
ISO point: Adjustment range:Complex start to fiducial point
Default: QRS onset to 28 msec
ST measurement point: Adjustment range:Fiducial point to complex end
Point default:QRS offset +80 msec
ST complex: Length: 892 msec (225 samples)
Frequency response:0.05 to 40 Hz
Update interval: 15 seconds, 1 normal beat required
Trend intervals: 1, 2, 4, 8, 12, 24 hours
Trend resolution: One data point every 30 seconds
ST level alarm adjustment range: 1 to 15 mm
ST event alarm duration: OFF, 15, 30, 45, 60 seconds
Alarm priority: Medium
Alarm auto set: Current value ±2 mm
Parameter display: Median frequency; Spectral Edge Frequency (SEF); Beta,
Alpha, Theta, and Delta powers; Burst Suppression Ratio
(BSR)
Channels: Up to 4 analog waveforms and parameter boxes, 4 channel
plot of power vs. frequency
Measurement range --
Median frequency:
Range: 0.5-30 Hz
Resolution:0.1 Hz
Accuracy:±10%
Measurement range --
Spectral Edge Frequency-SEF:
SEF Frequencies: 85, 90, 95, 97, 98%
Range: 0.5 - 30 Hz
Resolution:0.1 Hz
Accuracy:±10 %
Measurement range --
Total Power:
Range: 44 to100 dB
Resolution:1 dB
Accuracy:±10 %