B-38 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
Invasive Blood Pressure (IBP)
Degree of protection against
electric shock:
Type CF
Defibrillation protection: per EN 60601-2-30 (IEC 60601-2-30)
Measurement method: Resistive strain gauge transducer
Display resolution: 1 mmHg
Measurement range: –50 to +400 mmHg
Frequency ranges: DC to 8 Hz, DC to 16 Hz, and DC to 32 Hz
(user selectable)
Accuracy: ±1 mmHg or ±3 % exclusive of transducer
(whichever is greater)
Zero balance range: ±200 mmHg
Transducer specifications: Dräger provided transducers with a resistance of 200 to
3000 and an equivalent pressure sensitivity of 5V/V/
mmHg ±10 %
Degree of protection against electric
shock:
Type CF
Defibrillation protection: per IEC 60601-2-34