MONITORING S PECIFICATIONS
VF8 DELTA/DELTA XL/KAPPA B-41
Cardiac Output
Pulse Oximetry OxiSure SpO
2
via MultiMed
Parameter display: Cardiac output, blood temperature, Injectate
temperature
Measurement method: Thermodilution
Measurement range: Cardiac output:0.5 to 20 L/min
Blood temperature:25 to 43 °C (77 to 109 °F)
Injectate temperature:–5 to +30 °C (23 to 86 °F)
Accuracy: Cardiac output: ±5% (with 0° C injectate)
Injectate temperature: ±0.25 °C
Degree of protection against electric
shock:
Type CF
Defibrillation protection: Defibrillation-proof applied part per IEC 60601-1
Parameter display: Saturation (%SpO
2
), pulse rate
Measurement method: Transmission spectrophotometry
Measurement range: SpO
2
: 1 to 100 %
Pulse rate:30 to 250 1/bpm
Resolution: SpO
2
:1%
Pulse rate:1 b/min
Calibration range: 70 to 100 %
Display range: 0 to 100 %
Display update period: 2 seconds
Maximum hold from previous update: 30 seconds (in the event of artifact or other error)
Measurement accuracy, adult mode
(1, 2, 3, 5)
:
Saturation (%SpO
2
):
0 to 69% not specified
70 to 100% sensor-specific as follows:
Dräger :
OxiSure sensor - D ±2
Nellcor :
(4)
D-25/D-25L, D-20, I-20, N-25, OxiMAX MAX-A, OxiMAX MAX-AL,
OxiMAX MAX-P, OxiMAX MAX-N, OxiMAX MAX-I ±2
Nellcor :
DS100A ±3
Masimo:
(4)
LNOPADT, LNOPPED, LNOPNEO, LNOPNEO SS,LNOP-YI ±2
Masimo:
LNOP-DCI, LNOP-DCIP, NR125 ±2
EAR ±3.5
Pulse rate: ±3 beats/min or ±3% (whichever is greater)
Measurement accuracy and notes continued on next page.