D-41942-2012
Dräger Perseus® A500 Breathing System
Reprocessing Instructions
The following parts must be free of deposits:
– Incident ow grills
– Non-return valves (red and blue)
If required, remove deposits on the valve
plates of the non-return valves with a soft
cloth. If the incident ow grills are damaged,
contact DrägerService®.
* For machine cleaning, position each valve
so that the valve cage is at the bottom.
Visual check
Manual cleaning
and disinfection
Machine cleaning
and disinfection
Sterilization
Position components so that all interior spaces and surfaces are completely ushed and water can drain off freely.
This guideline does not replace the instruction for use.
*