EasyManuals Logo

Dräger Perseus A500 Manual

Dräger Perseus A500
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
Page 16 / 38
o condensation may block filters in the patient circuit (especially a filter at the inspiratory
port)
Therefore, we recommend that you:
o Do not to use a fresh gas flow less than 20% of the minute volume of the patient or
less than 1 L/min. Dealing with patients producing increased amounts of CO
2
(e.g.
having high fever) significantly raise the fresh gas flow (e.g. to >50% of the minute
volume).
o Use hose systems with water traps in the inspiratory and expiratory limb; the longer
the hoses, the better
o Do not use a filter at the inspiratory port of the breathing system
o Always use a mechanical filter, if available with HME at the Y-piece. Be aware that the
filters may get clogged and have to be exchanged earlier; therefore, set close alarm
limits for minute volume low and Paw high
o Perform a system test every 24 hours if possible and empty the piston membrane
before performing the system test (Apollo, Fabius family)
o Set an alarm limit for FiO
2
low with adequate buffer because the difference between
the set fresh gas flow and the FiO
2
will increase and the system will react far slower to
modifications of the O
2
setting
o Set an alarm limit for inspiratory CO
2
high to an appropriate value
o If required, update the default settings, refer to chapter 2.12
o Check the following parts for humidity more regularly (at least every 4 hours)
water traps in patient hoses (drain, if condensation is detected)
water trap at the gas bench - Waterlock2 (drain, if more than 50% is filled with
condensate)
filter (exchange if increased condensation is detected)
remaining capacity of CO
2
absorber (exchange at least when 2/3 has changed
its color to purple to reduce condensation in the breathing system;
consumption of CO
2
absorber will increase significantly when lowering the
fresh gas flow)
o The device has to be operated by a constant supervision of an operator with extensive
knowledge of a rebreathing system
The measures above are very important for keeping the installed base of ventilators in
your institution functional for the patients that need long-term ventilation.

Table of Contents

Other manuals for Dräger Perseus A500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Perseus A500 and is the answer not in the manual?

Dräger Perseus A500 Specifications

General IconGeneral
BrandDräger
ModelPerseus A500
CategoryMedical Equipment
LanguageEnglish

Related product manuals