© Drägerwerk AG & Co. KGaA, 2019
Savina Preparing for operation
Device check
2. Triggering the Apnoea alarm message
Activate ventilation mode CPAP/ASB
Set the PEEP key to 10 mbar (10 cmH
2
O).
Activate apnea ventilation.
Set the alarm limit T
Apnoea
to 15 s.
After the apnea alarm time T
Apnoea
elapses, the alarm message
Apnoea is displayed. The alarm signal sounds. Savina activates
apnea ventilation.
Steadily inflate and deflate the test lung to simulate spontaneous
breathing.
Press the Alarm Reset key.
Savina ventilates again in ventilation mode CPAP/ASB.
18
2
1
2
2
2