Savina Preparing for operation
Device check
3. Triggering the PEEP high alarm message
Activate ventilation mode IPPV
Set the alarm limit Paw to 100 mbar (100 cmH2O)
Put on sterile gloves and seal the flow sensor
The airway pressure will be released after the alarm limit Paw has
been reached. The alarm message PEEP high is displayed. The
red LED flashes. The alarm signal sounds.
Press the Alarm Reset key
19
3
3
2
3

Other manuals for Dräger Savina

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Savina and is the answer not in the manual?

Dräger Savina Specifications

General IconGeneral
FiO221 - 100 %
Ventilation ModesCPAP/ASB
DisplayColor
Power Supply100 – 240 V AC, 50/60 Hz
Patient RangeAdult, Pediatric, Neonatal
Tidal Volume Range20 to 2000 mL
Screen Size12.1 inch
Inspiratory Time0.1 - 10 s