Savina – Preparing for operation
Device check
4. Triggering the Airway pressure high alarm message
− On page Settings 1/1, deactivate the additional setting AutoFlow.
− On page Configuration 2/4, deactivate the ventilation function
Pmax
− On page Alarms 1/1, set the alarm limit Paw to a value less than
the current maximum airway pressure
The breath will be stopped after the alarm limit Paw has been
reached. The alarm message Airway pressure high is displayed.
The alarm signal sounds.
− Set the alarm limit Paw back to the hospital-specific value
− Press the Alarm Reset key