22 Dräger X-act 5000
Maintenance
4.2 Warnings
Warnings will be displayed together with the corresponding icon, and
if necessary, together with a written hint.
Possible warnings are:
mandatory leak test
needed activation strokes
needed desorption strokes
no measurement in areas of explosion hazard
low battery
invalid password
change SO
3
filter
PUMP <5°C/41°F START?
5 Maintenance
5.1 Maintenance table
5.2 Cleaning
The Dräger X-act 5000 can be cleaned using a damp cloth. Coarse
dust particles can be removed with a soft brush.
WARNING
To prevent ignition of flammable or combustible
atmospheres, disconnect the power supply before
maintaining the Dräger X-act 5000.
Work to be performed
every year
every
two years
Calibrate the flow rate. X
Have the Dräger X-act 5000 inspected by Dräger X
Change SO
3
filter.
X
1)
1) Or if indicated on the display
!