EasyManua.ls Logo

Dranetz-BMI ES210 - Programming; Factory Default; Parameters Overview

Default Icon
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
Parameters overview
Programming
All parameters may be displayed at any time. For modifications the jumper
on the backside of the device must be removed (not on position LOCK).
The following table shows all parameters with their adjustable ranges or
possible selections respectively. The black numbers give a cross-reference
to the appropriate diagram position on page 30.
Starting at the measurands display by pressing the key
P
you may
change to the menu level.
Afterwards you can select the desired menu item by pressing the key
P
shortly.
Use to enter the level where the desired parameter is displayed.
Pressing
P
shortly will force the selectable element to flash.
The flashing content may be modified using the keys or .
Press
P
for a longer time to leave the parameter or menu level.
All settings will remain non-volatile stored even in case of power-fail.
Hints:
First you have to set the system configuration and the transformer ra-
tios because further measurand selections, alarm limit settings etc. will
depend on them.
The programming may be modified via an optional extension module
as well.
Locking the configuration
Place the jumper in the LOCK position.
The configuration of all parameters is disabled.
LOCK
Factory Default
Brightness: (mid setting)
Limit value / S01: Off
Limit value / S02: Off
Transformer ratio: 1 : 1
Jumper: Not in the LOCK position
Connecting mode: 4 wire asymmetric load
Synchronizing
interval: 15 min.