1. Spread a clean white paper (white cloth) on the at table, point the
product against the white paper (white cloth), and then dry the product
with filtered dry compressed air (maximum pressure 3 bar). Until no
liquid is sprayed onto the white paper (white cloth), the product drying is
completed.
2. It can also be dried directly in a medical drying cabinet (or oven).
The recommended drying temperature is 80℃~120℃ and the time
should be 15~40 minutes.
Notes
a) The drying of product must be performed in a clean place.
b) The drying temperature should not exceed 138 °C;
c) The equipment used should be inspected and maintained regularly.
4.6 Inspection and maintenance
In this chapter, we only check the appearance of the product. After
inspection, if there is no problem, the handpiece should be immediately
reassembled, installing the sealing ring, LED, light guide, and cone head
in sequence to the handpiece, and then tighten the cone head clockwise.
4.6.1 Check the product. If there is still visible stain on the product
after cleaning/disinfection, the entire cleaning/disinfection process must
be repeated.
4.6.2 Check the product. If it is obviously damaged, smashed,
detached, corroded or bent, it must be scrapped and not allowed to
continue to be used.
4.6.3 Check the product. If the accessories are found to be damaged,
please replace it before use. And the new accessories for replacement
must be cleaned, disinfected and dried.
4.6.4 If the service time (number of times) of the product reaches the
specied service life (number of times), please replace it in time.
4.7 Packaging
Install the disinfected and dried product and quickly package it in a
medical sterilization bag (or special holder, sterile box).
Notes
a) The package used conforms to ISO 11607;
b) It can withstand high temperature of 138 °C and has sufficient
steam permeability;
c) The packaging environment and related tools must be cleaned
regularly to ensure cleanliness and prevent the introduction of