9
others buttons will go out.
3.Link the connector and the handpiece with the drainage device.
4.Step the pedal, the device would start self-cleaning. After that the
pedal could be loosen.
5.After cleaning for 30s, the device would stop self-cleaning. Or
u can stop by step the pedal again or press the automatic water supply
button.
6.After cleaning, press the automatically water supply button and
outside water button at the same time (1s) to exit the “Cleaning mode”
after buzzer beeps.
4. Cleaning, Disinfection and Sterilization
The cleaning, disinfection and sterilization of handpiece, tip, and
wrench (include torque wrench and Endo wrench) are as follow.
Unless otherwise stated, they will be hereinafter referred to as
“products”.
Warnings
The use of strong detergent and disinfectant (alkaline pH>9 or acid
pH <5) will reduce the life span of products. And in such cases, the
manufacturer takes no responsibility.
Do not clean the handpiece with ultrasound cleaning machine.
This device shall not be exposed to high temperature above 138℃.
Processing limit
The products have been designed for a large number of sterilization
cycles. The materials used in manufacture were selected accordingly.
However with every renewed preparation for use, thermal and chemical
stresses will result in ageing of the products. The maximum number of
sterilizations for handpiece is 600 times. For tips, it is 300 times. And for
wrench, it is 1000 times.
4.1 Initial processing
4.1.1 Processing principles
It is only possible to carry out effective sterilization after the
completion of eective cleaning and disinfection. Please ensure that, as
part of your responsibility for the sterility of products during use, only
sufficiently validated equipment and product-specific procedures are
used for cleaning/disinfection and sterilization, and that the validated
parameters are adhered to during every cycle.
Please also observe the applicable legal requirements in your country