Rinse handpieces through at least 3 times
using a 20-ml disposable pipette.
20 ml
Screw Scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times using a
20-ml disposable syringe.
20 ml
Remove all rinse adapters.
Comply with the cleaning agents' action times
specified by the manufacturer.
Intermediate rinsing
After the action time prescribed by the manufac-
turer:
Rinse all components under running water for
at least 1 minute (temperature < 20 °C).
Rinse handpieces through with water at least 3
times using a 20-ml disposable pipette.
20 ml
Screw scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times with
water using a 20-ml disposable pipette.
20 ml
Areas that are difficult to reach, e.g. the
Paro handpiece instrument chuck, should be
thoroughly rinsed (at least 5 x for 5 seconds
each) using an air and water syringe.
Usage
48 9000-615-28L02 1903V004
EN

Table of Contents

Other manuals for Durr Dental Vector Paro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental Vector Paro and is the answer not in the manual?

Durr Dental Vector Paro Specifications

General IconGeneral
BrandDurr Dental
ModelVector Paro
CategoryDental equipment
LanguageEnglish

Related product manuals