Rinse handpieces through at least 3 times
using a 20-ml disposable pipette.
❯
Screw Scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times using a
20-ml disposable syringe.
❯
Remove all rinse adapters.
❯
Comply with the cleaning agents' action times
specified by the manufacturer.
Intermediate rinsing
After the action time prescribed by the manufac-
turer:
❯
Rinse all components under running water for
at least 1 minute (temperature < 20 °C).
❯
Rinse handpieces through with water at least 3
times using a 20-ml disposable pipette.
❯
Screw scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times with
water using a 20-ml disposable pipette.
❯
Areas that are difficult to reach, e.g. the
Paro handpiece instrument chuck, should be
thoroughly rinsed (at least 5 x for 5 seconds
each) using an air and water syringe.