Screw Scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times using a
20-ml disposable syringe.
❯
Comply with the cleaning agents' action times
specified by the manufacturer.
Intermediate rinsing
After the action time prescribed by the manufac-
turer:
❯
Rinse all components under running water for
at least 1 minute (temperature < 20 °C).
❯
Rinse handpieces through with water at least 3
times using a 20-ml disposable pipette.
❯
Screw scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times with
water using a 20-ml disposable pipette.
❯
Areas that are difficult to reach, e.g. the
Paro handpiece instrument chuck, should be
thoroughly rinsed (at least 5 x for 5 seconds
each) using an air and water syringe.
Disinfection in an ultrasonic bath
NOTICE
Malfunctions in the handpiece due to
improper handling during cleaning or
disinfecting
❯
The handpieces of the Vector must
only be cleaned or disinfected in a suit-
able container in an ultrasonic bath.
❯
The handpieces must NOT be fully
submerged in fluid.
❯
Place the removable parts of the handpiece
(ring cover, rotary adaptor, scaler cover, light
conductor), instrument holders of the tool kits
(without instruments) and torque wrench in the
ultrasonic bath for the required action time, so
that all parts are covered.