The codes for the remaining digits in the alarm-operating mode are:
2nd Digit 3rd Digit 4th Digit 5th Digit
H = High Alarm A = Automatic reset d = Direct action n = Low alarm mask
L = Low Alarm n = Manual reset r = Reverse action blank = Not masked
For example, a display of I.HAr would indicate High alarm, automatic reset, reverse action.
2. Alarm 2
The indicator will automatically enter this parameter after the FUNC key is pressed to store the
Alarm 1 set point. To enter Alarm 2 from the normal operating mode, press FUNC five times.
Programming the Alarm 2 set point is the same as Alarm 1 above, except that the display
alternately shows 2.xxxx and the alarm value.
8.2.3 Alarm Reset Function
This function can be performed when the indicator is locked; must be in local mode. If the alarm
is configured as a latched alarm (manual reset), alarm status is maintained even after the alarm
condition stops.
Press RESET + either arrow ( or ) to reset both Alarm 1 and Alarm 2. The external contact, if
enabled, resets both alarms. The rear terminal connections are Dig In + and -. The external
contact works even if the indicator is in the remote mode.
8.2.4 Peak Hold Function
The following actions can be performed when the indicator is locked, and in either local or
remote mode.
1. Monitoring Peak High and Peak Low
By pressing while the measured value is displayed, it is possible to monitor the peak high value.
The decimal point at the right of the display will be lit steadily.
Press again to monitor the peak low value. The decimal point at the right of the display will now
blink on and off.
Press to redisplay the measured value (normal operating mode).
Press RESET + FUNC to reset the peak high/peak low values and to restart for a new peak
detection.
2. Hold-On Value
The external contact can be used to freeze input signal sampling, holding the last measured
value for use on the display, alarms, retransmission, etc. In this mode, the numerical value
flashes on the display.