CLEANING / MAINTENANCE
WARNING
Turn the Easy Vac Aspirator off and disconnect from any
external power source before cleaning.
Inline Hydrophobic Filter:
This filter cannot be cleaned. If it becomes contaminated or clogged,
it must be replaced.
To replace Inline Hydrophobic Filter:
1. Disconnect filter assembly from Suction Canister and remove
Suction Canister from the Aspirator.
2. Remove filter assembly from the Aspirator.
3. Dispose of filter.
4. Install new filter (Part #1690) and reconnect tubing and elbow.
Tubing is single patient use only. DO NOT clean.
Cleaning for Exterior Components:
1. Disconnect the Aspirator from power source and other
connections before cleaning.
2. Use a cloth or sponge dampened with soap and water.
3. Wipe dry with a clean cloth or paper towel.
4. Store the Aspirator in a clean area free from grease, oil, and other
sources of contamination.
RETURNS
Returned products require a Returned Goods Authorization (RGA)
number, contact Precision Medical, Inc. All returns must be packaged
in sealed containers to prevent damage. Precision Medical, Inc. will not
be responsible for goods damaged in transit. Refer to Precision Medical,
Inc. Return Policy available on the Internet; www.precisionmedical.com.
Manuals available on our Website; www.precisionmedical.com