59
Noise Level ≤15 μVp-p
Paper Speed
5mm/s, 6.25 mm/s, 10mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s,
Tolerance ≤±5%
Standard Sensitivity 10 mm/mV±0.2 mm/mV
Sensitivity
2.5 mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV,Tolerance ±5%
Internal calibration voltage
1 mV土5%
Polarization Voltage ± 500 mV
CMRR
>98 dB
2) Recorder
Recording Method Thermal dot array
Resolution
≥8 dots/mm(Vertical)
≥ 32 dots/mm (25 mm/s) ; ≥16 dots/m (50 mm/s) (Horizontal)
Recording Paper Roll Paper: 216mm*30M
Z fold Paper: 216mm*295mm*100P
3) Display
Display
12 inch color LCD screen, display resolution: : 800×600 (EM-1203)
10 inch color LCD screen, display resolution: : 800×480 (EM-1201&
EM-1202)
Display information
Paper Out, Lead Off Alarm, Operation Manual, Patient Information,
Battery Volume, Date and Time, Heart Rate, Working Mode, Paper
Speed, Gain, Filter
4) Others
Leads Standard 12 leads with defibrillator protection
Safety Level CF Type
Power Supply
AC: 100~240 V, 50/60Hz, 30 VA~100 VA
DC: 14.8 V/4400 mAh,built in rechargeable lithium battery
Fuse Specification
2-φ5x20 mm AC Time-delay Fuse T1 A/250 V
C.2 Dimension and Weight
Dimension: 41*30*10mm
Package Size 550*470*210mm