Appendix A1
A505-75-880 Issue B
Appendix A1 Correct use of Swagelok
connectors
Note: We recommend that you use a second spanner to hold the connector in position when you connect or
disconnect a Swagelok connector.
A1.1 Fit a Swagelok connector
1. Refer to Figure A1-1 detail A. Undo and remove the nut (4) from the Swagelok connector (1). Ensure that the
front (tapered) ferrule (2) and the rear ferrule (3) are correctly orientated as shown in detail A, then loosely
refit the nut (4) to the connector (1).
2. Refer to detail B. Insert the tube (5) through the nut (4) and into the Swagelok connector (1). Ensure that the
tube rests firmly on the shoulder inside the fitting, and that the nut (4) is finger tight.
3. Tighten the nut (4) until you cannot rotate the tube (5). If you cannot turn the tube because of how it is
installed, tighten the nut by 1/8th of a turn.
4. Refer to detail C. Mark the nut (4) at the six o’clock position.
5. Refer to detail D. Hold the body of the connector steady, then turn the nut (4) by 11/4 turns (to the nine o’clock
position) to fully tighten the connection.
A1.2 Reconnect a Swagelok connector
You can disconnect and reconnect a Swagelok connector many times and still obtain a correct leak-proof seal. Refer
to Figure A1-2 detail A which shows a Swagelok connector after you have disconnected it. Use the following procedure
to reconnect it:
1. Refer to detail B. Insert the tube (5) with the swaged ferrules (2, 3) into the Swagelok fitting (1), until the front
ferrule (2) is fully in the body of the fitting.
2. Refer to detail C. Tighten the nut (4) by hand.
3. Use a wrench or spanner to turn the nut (4) to its original position (you will feel an increase in resistance when