Installation
A505-75-880 Issue B
3 Installation
3.1 Safety
A suitably trained and supervised technician must install the Acoustic Enclosure.
If the GV80 pump has been in operation, switch it off and allow it to cool to a safe temperature before you
start installation.
Isolate the GV80 pump and other components in the process system so that they cannot be operated
accidentally.
Ensure that you connect, disconnect and tighten Swagelok connectors correctly: refer to Appendix A1.
When you refer to procedures in the GV80 instruction manual, ensure that you obey all of the appropriate
WARNINGs and CAUTIONs in the manual.
3.2 Unpack and inspect
Remove all packing materials and protective covers and check the Acoustic Enclosure. If any item is damaged, notify
your supplier and the carrier in writing within three days. State the Item Number of the Acoustic Enclosure, together
with your order number and your supplier’s invoice number. Do not use the Acoustic Enclosure if it is damaged.
Check that you have received the items listed in Table 1 (see page 4). If any item is missing, notify your supplier in
writing within three days.
If the Acoustic Enclosure is not to be used immediately, replace any packing and protective covers and store in
suitable conditions, as described in Section 5.1.
Obey the safety instructions given below and take note of appropriate precautions. If you do not,
you can cause injury to people and damage to equipment.