PTB600-MAN-V10.4 PTB 600 Instruction Manual
22
- KEY
This is the key to step the third menu level. Press or key to set the number recorded in the "
ACCESS KEY parameters at the third menu level, and press "F" + keys at the same time (it is not
necessary to confirm the choice by pressing the E key) to step to the third menu level. The acceptable
values are from 1 to 99: the default set value is 2. If you lost the access key remember that it is
possible to have the number by reading the register 13 (item 10.3)
3
rd
MENU LEVEL
MENU THAT CAN BE SELECTED BY PRESSING THE "F + " KEYS AT THE SAME TIME WHEN THE
KEY PARAMETER IS REACHED AT THE SECOND LEVEL AND WHEN THE SET VALUE
CORRESPONDS TO THE RECORDED ONE.
PRESS THE F KEY TO STEP THROUGH THE MENU.
PRESS THE “F + KEYS AT THE SAME TIME OR WAIT FOR ABOUT 20 SECONDS TO COME BACK
THE PRIMARY MENU
-
ACCESS KEY
Access key; numerical value from 1 to 99 that enables passing to the third parameter level. The default
value is 2
- BAUD RATE
Data transmission speed from the computer. Values are from 2400 to 19200 (default value is 9600).
- ADDRESS
Communication address. The value of this parameter is necessary to communicate from the computer to
many instruments. The admitted values are from 1 to 32 and once the value is set by using the or
keys it is necessary to confirm the choice by the E key
- S/N
Equipment serial number. It is set by the manufacturer and cannot be changed by the user.
- Board S/N
Serial number of the board. It is set by the manufacturer and cannot be changed by the user.
- MAX. SET.
Maximum value of the Set Point. It is set by the manufacturer and cannot be changed by the user.
- MIN. SET.
Minimum value of the Set Point. It is set by the manufacturer and cannot be changed by the user.
- WAIT
initial waiting procedure. If the value "0" is set, when it is started up, the calibrator immediately run to the
last set point value chosen after turning off. If the value "1" is set, when it is started up, the calibrator goes
on the waiting position and the SP flash. It is necessary to press any key in order to move it from the
waiting position and to choose the desired Set Point value. It is possible to set the WAIT value only by the
serial communication.
- REV. SOFTWARE
Internal software’s release number.
- SENSOR TYPE
It indicates the type of the internal probe.
- STAB:
It indicates the swinging value of the temperature, which has been set to see on the Display the symbol of
the oven ÷ steadiness. The symbol light on when the temperature is stable for over 6 minutes.