Setting Up
User Manual Biosen C_line Clinic / GP+
If the sensor is inserted incorrectly and the sensor flap
is forced down, the sensor will be damaged.
Close the sensor flap and press
the latch down on the closed side
until it audibly engages.
Figure 10 - Closing the sensor flap
Repeat this procedure for the second measuring channel if applicable.
Once the measuring channel is equipped correctly, the "Sensor
Replacement" window opens where the inserted sensors has to be
identified as "New".
After the sensor is inserted, the device needs some time for sensor
conditioning. This time is shown on the display and counts down. The
device is not ready for measurement during this period.
5.2.3.2 Sensor replacement
Chips sensors are subject to deterioration and must be replaced at
regular intervals. The sensor should be replaced at the latest after 60
days for glucose and after 50 days for lactate. Dummy sensors can be
used as long as there are no leaks in the flow system
It is possible to check the chip sensors with a sensor test solution or a
linearity solution (see section 16, page 54).
If adjustment of whole blood-/plasma glucose is
activated (see Info Menu "Info-4 Options") the sample type
must be set as Serum for measurement of sensor test solution or
linearity solution.
Only Replace sensors when then device is connected to the mains
supply.
Risk of infection, wear suitable gloves!