Info Menu
User Manual Biosen C_line Clinic / GP+
32
32
8. Info Menu
The Info Menu contains 5 windows, reachable via the arrow buttons,
which contain important information about the device in a brief format.
"Info-1 Sensor"
Type: Sensor type (Glucose/Lactate/Dummy)
Active: Measuring channel activated (yes/no)
Calibration: Value of the last calibration
Age: Age of current sensor
"Info-2 Counter"
Current running counter numbers.
"Info-3 Statistic"
Today: Number of new data
Total: Number of total data
"Info-4 Extras"
Adaptation: Measured values can be adapted to other glucose or
lactate measuring methods. The measured values are
converted by using the entered equation and displayed.
Modifications can only be made by service staff or with a
service card. The default setting is y=1.00x+0; no
adaptation of the measured value.
H-Factor: Factor for adjustment of whole blood glucose and
plasma glucose. If the adjustment is activated (H-Factor
1,11) and the sample type hemolysate is set the
displayed value is increased by 11%.
Modifications can only be made by service staff or with a
service card. The default setting is 1,00, hence the
selected sample type has no influence on the displayed
measurement result.
"Info-5 Device"
Model number
Serial number
Software version HMIC/PIC

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EKF Diagnostics Biosen C-Line GP+ and is the answer not in the manual?

EKF Diagnostics Biosen C-Line GP+ Specifications

General IconGeneral
BrandEKF Diagnostics
ModelBiosen C-Line GP+
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals