EasyManua.ls Logo

EKF Diagnostics Biosen C-Line GP+ - Warning limits; Samples; Control 1; Control 2

Default Icon
123 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Device menu
User Manual Biosen C_line Clinic / GP+
35
If adjustment of whole blood glucose/plasma glucose is
activated it is mandatory to select the correct sample type, as this has an
influence on displayed results.
9.3.5 Sample preselection
Activates / deactivates the sample preselection.
9.3.6 Tray identification
This menu item is not available.
9.4 Warning limits
9.4.1 Samples
Setting of warning limits for sample measurements for both measuring
channels. Possible settings are:
Glucose 0,5mmol/l (9mg/dl ) 50mmol/l (900mg/dl )
Lactate 0,5mmol/l (5mg/dl ) 40mmol/l (360mg/dl )
9.4.2 Control 1
Setting of warning limits for special position C1. The defined upper and
lower limit must agree with the permitted limits of the used control
material.
9.4.3 Control 2
Setting of warning limits for special position C2. The defined upper and
lower limit must agree with the permitted limits of the used control
material.
9.4.4 Range violation
These settings control the action taken, when defined limits are
exceeded.
"Controls continue measurement?"
If limits are exceeded in controls C1 or C2, measurement of the
samples can be aborted or continued.

Table of Contents

Related product manuals