EasyManua.ls Logo

EKF Diagnostics DiaSpect Tm - 9. Troubleshooting Guide; Error E01: Calibration Lost; Error E02: Sensor Read Error; Error E03: Self-check Failed

EKF Diagnostics DiaSpect Tm
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
DiaSpect Tm User Manual
9. Troubleshooting Guide
Symptom
Possible Cause
Correction
Unexpectedly
high / low result
Improper sample
Repeat the sampling. Make sure that
the sampling is done correctly.
See 7.1 to 7.3 for more information.
Error E01
Calibration lost
Contact Technical Support at
1-800-531-5535.
Error E02
Sensor read error
Repeat measurement with the same
cuvette. If error persists, contact
Technical Support at 1-800-531-5535.
Error E03
Self-check failed
E03 may be displayed if a filled cuvette
is left in the cuvette holder, or was
removed too slowly. In order to reset
the self-check function, press down on
the empty cuvette holder. The screen
should display “---” and a “√”.
If error persists, contact Technical
Support at 1-800-531-5535.
Error E04
Light source too dark
Remove cuvette from cuvette holder.
Press cuvette holder several times until
the screen reads “---and a √”
appears. If error persists, contact
Technical Support at 1-800-531-5535.
Error E05
Light source too bright
Remove cuvette from cuvette holder.
Press cuvette holder several times until
the screen reads “---and a √”
appears. If error persists, contact
Technical Support at 1-800-531-5535.
Error E07
Battery too low to
perform measurements
Recharge the battery.
Display blank,
measuring not
possible
Battery completely
discharged
To recharge the battery, connect with a
power outlet or computer and charge for a
minimum of 4 hours.
If recharging fails, contact Technical
Support at 1-800-531-5535.

Other manuals for EKF Diagnostics DiaSpect Tm

Related product manuals