3
DiaSpect Tm User Manual
3. The DiaSpect Tm system
3.1 DiaSpect Tm analyzer
Upon delivery, open the carton on a stable surface, remove the instrument and
the accessories, and check that all the components are included and undamaged.
A comprehensive list (with item numbers) of all parts for the DiaSpect Tm system
can be found in sections 15 and 16.
The DiaSpect Tm analyzer can be stored at 0 to 50 °C (32 to 122 °F).
Temperatures of -30 to 70 °C (-22 to 158 °F) are temporarily permitted during
transport (24 hours max.).
The operating temperature of the instrument is 10 to 42 °C (50 to 107 °F). Allow
the analyzer to reach ambient temperature before use.
1. DiaSpect Tm
analyzer
2. User Manual
3. Power supply, adapter
plug and USB cable
3.2 DiaSpect Tm Cuvettes
Cuvettes are ready for use upon removal from the package. A sample volume of
10 µL is required to ensure proper filling of the DiaSpect Tm Cuvette. The cuvette
serves as sample collector and measuring cuvette at the same time. The blood
sample is drawn into the cavity by capillary force.