EasyManua.ls Logo

EKF Diagnostics Stanbio HemoPoint H2 - Control Cuvette

EKF Diagnostics Stanbio HemoPoint H2
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sampling and testing procedure
HemoPoint
®
H2 User Manual
28
Always carry out a blank reading whenever you have removed
the cuvette holder (i.e., cleaning, replacing with a new cuvette
holder).
A blank reading can be performed as follows:
1. Open the cuvette holder completely.
The text in the display reads “Add Cuvette”.
2. Close the cuvette holder, without inserting a cuvette.
In this case the analyzer does not start testing. After approx. 2 –
3 seconds the display will read “Open Holder”. If the instrument
does not show “Open Holder”, there is possibly a malfunction.
Please consult Section 7 “Troubleshooting”.
4.3.3 Control cuvette
The control cuvette supplied with the analyzer can be used for a
simple check of the analyzer’s calibration.
Illustration 23: Control cuvette.
The Hgb value and the permitted deviation of the control cuvette
are stated on the storage box label. All analyzers have a specific
calibration due to the tolerances in the mechanical and electronic
components. The control cuvette is only calibrated for the
TIP
TIP

Table of Contents

Related product manuals