DUO - User Manual Pg 16
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Eko Electronic Stethoscope System
DUO is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The user of DUO can help prevent electromagnetic interference by maintaining a minimum

Rated Maximum Output
PowerofTransmier(W)
SeparationDistanceAccordingtoFrequencyofTransmier(m)
150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 12 12 23


NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2:
6.1 Electrical Safety