EKO INSTRUMENTS CO., LTD. I-V Curve Checker MP-11 Instruction Manual Ver.7
Pg. 37
8-2. Starting MP-11
Always switch ON the MP-11 power FIRST then connect the PV module/array.
Always disconnect the PV module/array FIRST then Switch OFF the MP-11 power.
1. Start Up
Insert the AC adaptor or Battery Receptacle Box plug in the MP-11 Power Supply connector.
To turn MP-11 on, press the POWER button on the operation panel.
2. Start Screen
When the MP-11 Main Unit is turned ON, the following screen appears on the LCD display.
Firmware version number
Battery Status Gauge
When using Battery for power supply, the Battery Status Gauge is
displayed as the Main Unit is powered ON.
NOTE: The Battery Status Gauge is only for your reference; there
may be some differences for the actual remaining battery power
3. [Powering on Sensor Unit] Message
On MP-11, there is a function to synchronize the time between the Main Unit and Sensor Unit when the Main
Unit is powered ON. If the Sensor Unit is not connected to the Main Unit or Sensor Unit power is not turned
ON even though it is connected to the Main Unit, the “Powering on Sensor unit.” Warning message appears
on the screen.
Warning Message Powering on Sensor unit.
If the Sensor Unit is not used, press any button to remove the warning
message.
The Main Unit cannot be powered OFF with this warning message
displayed; press any key and remove this message before powering
OFF.
For more details of waning messages, refer to [9-2. Error and Warning
Messages]
Using the Sensor Unit to take measurements:
Connect the Main Unit and the Sensor Unit with modular cable; turn ON the Sensor Unit power first then
turn ON the Main Unit power. Always follow this order when the Sensor Unit is used.
After the time synchronization between the Sensor Unit and the Main Unit, the SU is indicated at right
top on the Main Menu screen.
PORTABLE I-V CHECKER
MP-11
Ver.04.001
EKO INSTRUMENTS
Start Screen
PORTABLE I-V CHECKER
MP-11
Ver.1.0.0
EKO INSTRUMENTS
Powering on
Sensor unit.
Powering on Sensor Unit
Warning Message