EKO INSTRUMENTS CO., LTD. I-V Curve Checker MP-11 Instruction Manual Ver.7
Pg. 42
Parameter Editing Steps
USE SENSOR UNIT
MAIN UNIT ONLY
[Voltage Range] Select a voltage range
[Current Range] Select a current range
[Sensor] Select a radiation sensor
(Pyranometer: PYR/Ref. Cell: REF)
[Cal] Enter Pyranometer/Ref. Cell sensitivity
When Sensor: PYR is selected for the Sensor,
enter the sensitivity value of the pyranometer.
For the internal pyranometer, the sensitivity is stated on
the back of the Sensor Unit. For the external
pyranometer, use the sensitivity stated on the
pyranometer.
NOTE: When external pyranometer is used,
connect the pyranometer to PYR terminal and set
the switch to the EXT side.
When Reference Cell: REF is selected for the
Sensor, the entering unit for the sensitivity will be in
[mV/kWm
2
] and setting value will be determined by
the following calculation:
[Reference Cell Current Value]
(mA/kWm^2) × 0.1(Ω)
For Example, when connecting a reference cell which
flows 125mA at 1kW/m2, it will generate the following
amount of voltage:
125mA×0.1Ω 12.5mV
Therefore, enter 12.5 [mV/kWm
2
].
NOTE: When reference cell is used, connect it to
the REF terminal.
NOTE: Reference cell which flows more than
500mA cannot be connected to REF terminal.
Select whether to use fixed values for the irradiance
and temperature.
Selected [YES]:
Enter the irradiance value in [Irradiance]
Enter temperature value in [Module temp]
Selected [NO]:
[Voltage Range] Select the voltage range
(100V, 600V, 1000V, AUTO)
[Current Range] Select the current range
(2A, 10A, 30A, AUTO)
Irradiance & Temp Input Options
01 PARAMETER EDIT
NO
YES
Set Irr & Temp ?
Range Option
Name EKOTEST1
Voltage Range AUTO
Current Range AUTO
01 PARAMETER EDIT
Irradiance: 1000.00 W/m
2
Module temp.: 25.0
Irradiance & Temp Fixed Value Input
01 PARAMETER EDIT
Range & PYR Calibration Value Setting
Name EKOTEST1
Voltage Range AUTO
Current Range AUTO
Sensor PYR
Cal 46.00 mV/kWm
2
01 PARAMETER 1/2