EKO INSTRUMENTS CO., LTD. I-V Curve Checker MP-11 Instruction Manual Ver.7
Pg. 78
2. Individual Conversion Data File(CSV Conversion Method)
By selecting the individual conversion, following data format is generated.
Name
String-3
Date
2014/3/9
Time
12:30:42
Temperature[degC]
31.752979
Voltage Range[V]
600
Current Range[A]
10
Sampling Pitch[uS]
10
Sensor
Pyranometer
Calibration Constant (Er cal)[mV/kW/m^2]
49.94
PV Device Area[m^2]
1.6421
Current Temp. Coef.[A/degC]
0.00463
Voltage Temp. Coef.[V/degC]
-0.135
Series Resistance(Rs)[ohm]
0
Curve Correction Factor (Kappa)[ohm/degC]
0
Series Modules
11
Parallel Modules
1
Solar Irradiance (Er)[W/m^2]
1084.70803
PV Device Temp.[degC]
57.502277
Isc[A]
9.822796
Voc[V]
376.202089
Pm[W]
2523.12274
Ipm[A]
8.915196
Vpm[V]
283.013728
Eta[%]
12.87755
FF
0.682782
Er (stc)[W/m^2]
1000
T (stc)[degC]
25
Isc (stc)[A]
8.95452
Voc (stc)[V]
418.278064
Pm (stc)[W]
2656.561412
Ipm (stc)[A]
8.131098
Vpm (stc)[V]
326.71618
Eta (stc)[%]
14.707118
FF(stc)
0.709271
Voltage[V]
Current[A]
0
9.823
0.12
9.829
:
:
:
:
*) 402 pairs of voltage and current are stored in CSV file.
*) The first pair shows data of Isc point. The last pair shows data of Voc point.