EKO INSTRUMENTS CO., LTD. I-V Curve Checker MP-11 Instruction Manual Ver.7
Table 11-1. Specification - Continued
I-V curve graph, P-V curve graph, Derivation Graph
・ I-V curve graph partial enlargement function
・ Open Circuit Voltage: Voc, Short Circuit Current: Isc, Max. Output Power: Pm
・ Max. Output Power Voltage: Vpm, Max. Output Power Current: Ipm, Fill Factor: FF,
Power Generation Efficiency: η
・ STC Conversion Values (Voc, Isc, Pm)
PV parameter setting function (α, β, κ)
・ Complying to IEC 60891 and JIS C8914/8919/8940
・ Conversion by measured Irradiance, Module temperature
Note) No conversion to AM1.5 by spectrum
PV Module Parameter Settings: (α, β, κ, Rs, module size)
Parameter setting save capacity: 70 settings
Automatically calculates STC conversion from the irradiance measured by sensor unit
pyranometer/reference cell and module temperature and display on graph.
・ Use fixed irradiance and module temperature values setup prior to measurement when
sensor unit is not connected.
Within 5 seconds (Time between pressing Measure button and
measurement result is displayed)
・ Min. Measurement Interval: 30 sec
(Time between last measurement to the next measurement)
・ Waiting time till the next measurement is indicated in 2 steps
・ Interval Time: 1 to 60minutes
(Time between measurement start to the next measurement start)
Setting Resolution: 1 minute
・ Measurement Setting Time Range: 00:00~23:59
*) Measurement beyond midnight cannot be set.
Measurement
Data Save
Function
・ Saved Data Quantity: 300 data
(Data is overwritten when it exceeds 300)
・ Search saved data、display graph & parameter
・ Display graph with Irradiance vs. Voc/Isc/Pm/FF from saved data
・ Saved data deletion
In Manual Mode, measure and display the Voc, irradiance, module
temperature and ambient temperature (when connected with Sensor
Unit) every 2 seconds.
4mmφ Test Lead Terminal (+, -), FG Terminal
Measurement accuracy within 1.5% of Full Scale
Integrated Pyranometer:
External Pyranometer Input:
- Applicable with EKO MS-602/802, ML-020VM, ML-01
- Pyranometer mode changed by switch on Sensor Unit rear panel
*) Setup sensitivity as parameter for using external pyranometer
< 500 mA:
Within 2.0% of Full Scale
-50℃ to +100℃ (T-type)
2ch: Module temp. & ambient temp.
Disconnect Sensor Unit from Main Unit, log the sensor data then upload the
sensor data by I-V measurement timing as Sensor Unit connected after the
measurements to execute the STC conversion.