EKO INSTRUMENTS CO., LTD. I-V Curve Checker MP-11 Instruction Manual Ver.7
Pg. 88
Table 11-1. Specification - Continued
Items
Specifications
Graph Display
I-V curve graph, P-V curve graph, Derivation Graph
I-V curve graph partial enlargement function
Measurement
Parameters
Open Circuit Voltage: Voc, Short Circuit Current: Isc, Max. Output Power: Pm
Max. Output Power Voltage: Vpm, Max. Output Power Current: Ipm, Fill Factor: FF,
Power Generation Efficiency: η
STC Conversion Values (Voc, Isc, Pm)
STC Conversion
Functions
PV parameter setting function (α, β, κ)
Complying to IEC 60891 and JIS C8914/8919/8940
Conversion by measured Irradiance, Module temperature
Note) No conversion to AM1.5 by spectrum
PV Module Parameter Settings: (α, β, κ, Rs, module size)
Parameter setting save capacity: 70 settings
Automatically calculates STC conversion from the irradiance measured by sensor unit
pyranometer/reference cell and module temperature and display on graph.
Use fixed irradiance and module temperature values setup prior to measurement when
sensor unit is not connected.
PV Measurement
Functions
Measurement
Time
Within 5 seconds (Time between pressing Measure button and
measurement result is displayed)
Manual
Measurement
Min. Measurement Interval: 30 sec
(Time between last measurement to the next measurement)
Waiting time till the next measurement is indicated in 2 steps
Auto
Measurement
Interval Time: 1 to 60minutes
(Time between measurement start to the next measurement start)
Setting Resolution: 1 minute
Measurement Setting Time Range: 00:0023:59
*) Measurement beyond midnight cannot be set.
Measurement
Data Save
Function
Saved Data Quantity: 300 data
(Data is overwritten when it exceeds 300)
Search saved datadisplay graph & parameter
Display graph with Irradiance vs. Voc/Isc/Pm/FF from saved data
Saved data deletion
Monitor
Function
In Manual Mode, measure and display the Voc, irradiance, module
temperature and ambient temperature (when connected with Sensor
Unit) every 2 seconds.
PV Input Terminal
4mmφ Test Lead Terminal (+, -), FG Terminal
Other Sensor
Functions
Sensor
Unit
Pyranometer Input::
Measurement accuracy within 1.5% of Full Scale
Integrated Pyranometer:
External Pyranometer Input:
< 1.5kW/m
2
< 100mV
- Applicable with EKO MS-602/802, ML-020VM, ML-01
- Pyranometer mode changed by switch on Sensor Unit rear panel
*) Setup sensitivity as parameter for using external pyranometer
Reference Cell Input:
< 500 mA:
Within 2.0% of Full Scale
Thermopile Input:
-50 to +100 (T-type)
2ch: Module temp. & ambient temp.
Remote
Sensing
Functions
Disconnect Sensor Unit from Main Unit, log the sensor data then upload the
sensor data by I-V measurement timing as Sensor Unit connected after the
measurements to execute the STC conversion.