What to do if the Eletta V Series meter is not giving an alarm?
- AalexanderronaldAug 6, 2025
If your Eletta Measuring Instruments meter isn't giving an alarm, start by checking the power supply connection. Then, ensure the monitor is correctly mounted, aligning the flow direction arrow on the pipe section with the actual flow for GL- and FA-models, or by checking the flow direction selector inside the monitor. Also, verify there is sufficient flow in the pipe to create the needed differential pressure. Confirm you have the correct orifice plate for the application by checking the stamped values. Finally, if using compression couplings, ensure the inside diameter is adequate to avoid a 'nozzle' effect.