Product Data Sheet
00813-0100-4530, Rev DA
December 2010
21
Rosemount 5300 Series
Display and Configuration
Integral Display
(Options Code M1)
The integral digital display can toggle between: level, distance, volume, internal temperature,
interface distance, interface level, peak amplitudes, interface thickness, percentage of range, analog
current out
Note: The display cannot be used for configuration purposes
Remote Display Data can be read from the optional integral display or remotely using the Rosemount 751 Field Signal
Indicator for 4-20 mA / HART (see Product Data Sheet, Document No. 00813-0100-4378), or the
Rosemount 752 Remote Indicator for F
OUNDATION™ fieldbus (see Product Data Sheet,
Document No. 00813-0100-4377)
Configuration Tools HART: Rosemount Radar Master, Emerson Field Communicator, AMS Device Manager, or any other
DD (Device Description) compatible host system
F
OUNDATION™ fieldbus: Rosemount Radar Master, Emerson Field Communicator, DeltaV, or any
other DD (Device Description) compatible host system
DTM (compliant with version 1.2 of the FDT/DTM specification) supporting configuration in for
instance Yokogawa Fieldmate/PRM, E+H™ FieldCare, and PactWare™
Output Units Level, Interface and Distance: ft, inches, m, cm, or mm
Level Rate: ft/s, m/s, in./min, m/h
Volume: ft
3
, inch
3
, US gals, Imp gals, barrels, yd
3
, m
3
, or liters
Temperature: °F and °C
Output Variables
Diagnostics Transmitter diagnostics with alerts include hardware and software errors, electronics temperature,
probe missing, and invalid measurement and configuration error diagnostics. In addition to this, echo
curve and variable logging including signal strength facilitate easy on-line troubleshooting
Advanced Diagnostics Signal Quality Metrics monitors the degradation of level signal relative to excess noise.
This can schedule the cleaning of the probe or to detect and monitor turbulence, boiling, foam, and
emulsions. It can be ordered by choosing the D01 or DA1 option
Damping 0-60 s (2 s, default value)
5301 5302 5303 PV, SV, TV, QV
Level XXX X
Distance to Level (Ullage) XXX X
Level Rate XXX X
Signal Strength XXX X
Volume XXX X
Internal Temperature XXX X
Interface Level (X)
(1)
(1) Interface measurement only for fully submerged probe, see page 25
XX
Interface Distance (X)
(1)
XX
Interface Level Rate (X)
(1)
XX
Interface Signal Strength (X)
(1)
XX
Upper Layer Thickness XX X
Lower Volume (X)
(1)
XX
Upper Volume (X)
(1)
XX
Signal Quality XXX(X)
(2)
(2) Not available as primary variable
Surface/Noise Margin XXX(X)
(2)
Vapor DC X(X)
(2)
Analog Output Current
(3)
(3) LCD variable only. Not available for FOUNDATION fieldbus, Modbus Signal Output, or
for HART units in fixed current mode
XXX
% of Range
(4)
(4) LCD variable only. Not available for FOUNDATION fieldbus Signal Output
XXX